Magnetoresistive Milligauss Meter MR3
The MR3 Magnetoresistive Milligauss Meter measures and displays the three vectorcomponents (X, Y, Z) as well as the magnitude (M) of the static magnetic field. Update is two times per second. The three sensors are in a 25 mm (1”) cube at the end of a flexible 100 cm (40”) cable. It can be powered by 4 AA batteries (included) or a Worldwide AC adaptor (included). It can collect and store data (X, Y, Z and M) for later download or stream data live via USB connector. Both the instantaneous values of X, Y, Z and M, and their peak values, are displayed on the meter screen. “Peak” values are the highest values seen since the peak reset button was pressed. Maximum error of sensitivity is +/- 0.5 % of the reading, and the maximum error of the offset is +/- 0.50 milligauss. Range is +/- 1999.99 milligauss with resolution 0.01 milligauss (or 1 nT) over the full range, displayed simultaneously for X, Y, Z and M. There is a “Zero” button which subtracts the instantaneous field for relative (tare) measurements, if required.