Product type
Sensor
Measuring range
0
100 %
Device type
Portable, Stationary
Gasflow
Flow by diffusion, Forced flow
Technology
Electrochemical
Approvals
Ja
Output signal
9-13 mV in air
Features
Expected operating life > 60 month in normal operation, very low cross sensitivity and fast responding electrochemical sensor
Product description
Electrochemical sensor for medical applications
Oxygen (0,...,100%)
mV output signall - 9-13 mV in air
Repsonse time (T90) less than 7 seconds
Lifetime more than 4 years
RoHS compliant
KEY FEATURES
Lead free, RoHS compliant, long-life expectancy.
All characteristics are based on conditions at 25°C, 50% RH and 1013 hPa and gas flow 2.5 L/min.
Measurement Range: 0 to 100 Vol.% O2
Expected Operating Life: 4 years @ ambient air, depending on application
Nominal Sensor Life: > 2, 000, 000 Vol.% O2 h
Electrical Connector: 3-pin Molex
Mechanical Mating Connector: fits for M16x1 DIN 13 or 5/8-24 UNEF
Initial Output Signal Range: 9 - 13 mV @ dry ambient air
Response Time t90: < 7 s
Signal Drift (long term): < 0.1%/month of sensor output signal @ dry ambient air
Signal Drift (short term): < 0.1 Vol.%O2/day @ dry ambient air, constant environmental /
measurement conditions
Static Temperature Error: < ±1.5% @ 20-40 °C
< ± 2% @ 10-20 °C
< ±10% @ 0-10 °C / 40-50 °C
Operating Temperature: 0 - 40 °C; intermittent 40 - 50 °C
Ambient Pressure Range: 500 to 1250 mbar
Zero Offset Equivalents: < 0.3 Vol.%O2 @ 100 Vol.% N2 applied for 5 min
Linearity Error: < 3 % @ 100 Vol.% O2 applied for 5 min
Influence of Humidity: - 0.03 % rel. O2 reading per % RH
Temperature Compensation: NTC on sensor PCB