-Quick and sensitive test
-Zero alteration of container features
-Vacuum
-Fast, reliable and repeatable results
-Highly functional, intuitive HMI
-Cost-effective solution
-System autodiagnostics available
Technical Features
-Container Application: Ready Meals, Food, Stick Packs, Doy Pack, Beverage
-Products: Filled, Empty, Closed
-Container Dimensions: From 5 x 30 x 5 mm (min) to 86 x 86 x 250 mm (max)
-Speed: Up to 20 cpm
-Technology: CCIT
-Inspection features: Non-Invasive, Non-Destructive CCIT based on Vacuum Decay Method
-Inspection Capabilities: Microleaks detection
Additional Benefits
-Compact and robust design
-Maintenance-free
-Quick change over
-Low power consumption
-Storage, maintenance, and download of historical data (production, raw data, events, alarms)
-HMI Real Time display of Testing Cycle diagram
-Computerised system is designed to comply with FDA 21 CFR Part 11 and EU Annex 11
-Validation Package guarantees complete and efficient regulatory compliance
Technology
-Container Closure Integrity Testing is a non-destructive measurement technology based on Vacuum Decay Method.
-Measurement system comprises applying a pressure differential into an airtight testing group enclosing the container.
-The test objective is to detect container leakages by measuring the reached pressure level as well as the pressure change over test time.