Trace Hydrogen Analyzer
Features:
1.Latest trace hydrogen tester;
2.Big LCD screen, concise menu;
3.Connecting to the printer to print the hydrogen concentration in a certain termina;
4.Automatic data storage in a certain intervatime, and browsing data at any time;
5.Two-way communication by connected to a computer via the standard RS232 interface;
6.Setting upper and lower alarm values arbitrarily during the measurement range;.
7.Calibrating by standard gas;
8.0-10mA or 4-20mA electricacurrent output;
Industries Served:
To measure the trace hydrogen concentration in the fields of nitrogen separation, chemical process, magnetic materials such as high temperature sintering
furnace gas protection, various mixed gases, etc.
Range:
0-50ppm, 0-500ppm, 0-5000ppm H2
Indicating error:
±2% of full scale range
Repeatability:
2%
Response time:
T90<40s
Contact capacity:
250VAC/1A
Signal output::
4-20mA DC or 0-10mA DC
Power supply:
220±10% VAC, 50Hz±5%
Environmental temperature:
0℃-+45℃
Sample gas flow speed:
300-400mL/mim
Sample gas pressure:
≤0.2MPa
Appearance dimension:
133mm×133mm×250mm(W×L×D)
Weight:
2.5kg