Introduction
Infrared Syngas Analysis System Gasboard-9021 is a continuously monitoring system which consist of sampling unit, pretreatment unit, gas analysis unit and PLC controlling unit, it can provide unattended real time monitoring of CO, CO2, CH4, H2 , O2 , CnHm, C2H2 , C2H4 or any of their combination for various applications.
Features
• PCT approved product based on IR & TCD technologies to measure syngas composition. (EU and US Authorization No.: EP2796856 , US9857323)
• Replacement to gas chromatography, mass spectrometry.
• Patented NDIR and TCD technology, including ECD technology for reliable measurement of eight gases.
• High-selectivity CH4 gas sensor, no interference from other hydrocarbons.
• Fully compensated H2 sensor, no interference from CO, CO2, CH4 based on intelligent correction
• Integrated with gas sampling &conditioning part, analysis part and controlling part, suitable for continuous unattended real time monitoring.
Specifications
Gas Measured: CO, CO2 , CH4, CnHm, C2H2, C2H4 by NDIR sensor; H2 by thermal conductivity detector; O2 by electrochemical sensor (Paramagnetic O2 sensor on request)
Measurement Range: CO/CO2/CH4/H2:(0~100)%; O2:(0~25)%; C2H2/C2H4/CnHm:(0~10)%(optional)
Accuracy: CO/CO2/CH4/C2H2/C2H4/CnHm:±1%F.S.; O2/H2:±2%F.S.
Resolution: 0.01%
Repeatability: ≤1%F.S.
Response Time: T90<15s (NDIR)
Optimal flow Rate: (0.7~1.2) L/min
Inlet Gas Pressure: (2~50) kPa
Gas Condition: No Tar, No Dust and No Water
Output Communication: 4-20mA
Power supply: AC220V,50Hz
Weight: 250kg
Size: 1800mm(L) *700mm(W) *600mm(H)