NO sensor Gasboard-2102
micro-flow NDIRRS-485RS-232

NO sensor - Gasboard-2102 - Cubic Instruments (Wuhan) Ltd. - micro-flow NDIR / RS-485 / RS-232
NO sensor - Gasboard-2102 - Cubic Instruments (Wuhan) Ltd. - micro-flow NDIR / RS-485 / RS-232
Add to favorites
Compare this product
 

Characteristics

Gas type
NO
Technology
micro-flow NDIR
Output
RS-485, RS-232
Other characteristics
miniature, maintenance-free

Description

Introduction Gasboard-2102 is derived from the Gasboard-2000 series with the same advantage of micro flow NDIR gas bench, but focuses on low concentration NO detection, which is a high precision NDIR NO gas bench with compact design and very suitable for gas analyzer manufacturer to integrate .The lowest detection concentration can reach 0~200 ppm Features • Advanced Micro flow NDIR working principle for low range NO detection • Dual chamber for high precision • Modular design for easy integration • Automatic temperature and pressure compensation Specification Measurement:NO NO:0~200~5000ppm Measurement range can be customized Linear Error:<±2% FS Repeatability:≤2%FS Zero/Span drift:<2.5% FS/ 7days Applications Motor vehicle exhaust emission test Atmospheric/exhaust gas environmental testing