Designation: CO, CO2, O2, H2O
With pressure-compensated paramagnetic oxygen sensor and P2O5 - Analyzer unit for residual moisture measuring.
The gas analysis system described hereafter is designed for set-up in safe area and under air conditioned, in-door conditions. It provides continuous and simultaneous concentration measurement of the following compounds, which is supplied from one source:
• Traces of Carbon Dioxide (CO2) 0-500ppmv
• Traces of Carbon Monoxide (CO) 0-50ppmv
• Purity of Oxygen (O2) 95-100Vol%
• Traces of Humidity (H2O) 0-2000ppmv - AQUATRACE® 19" inch rack
• Background gas: oxygen
This system contains all necessary equipment for sampling, conditioning and analysis of above listed gases.
The following measuring methods are used to determine the concentration of each individual compound:
• CO ppm V GfX infrared gas analyzer
• CO2 ppm v GfX infrared gas analyzer
• O2 Vol% Pm paramagnetic gas analyzer (with atmospheric pressure compensation)
• H2O ppm V P2O5 trace moisture analyzer 0-2000 ppm V