Reference instrument for the calibration of differential pressure measuring instruments under static pressure
Primary pressure standard for calibration laboratories and factory, for the testing, adjustment and calibration of differential pressure instruments
Complete stand-alone system, also suitable for on-site use
Proven primary pressure standard
Deadweight Tester / Pressure Balances are the most accurate instrument available on the market for the calibration of electronic or mechanical pressure measuring instruments. The direct measurement of the pressure (p = F / A), as well as the use of high-quality materials enable a very small measurement uncertainty, in conjunction with an excellent long-term stability of 5 years (recommended in accordance with German Calibration Service DKD/DAkkS).
The deadweight tester / pressure balance has therefore been used for years in factory and calibration laboratories in national institutes, research laboratories and in industry.
Stand-alone operation
Due to its integrated pressure generation and the pure mechanical measuring principle, the model LR-Cal CPB5600-DP is ideal for on-site use for maintenance and service.
Basic principle
Pressure is defined as the quotient of force and area. The core component of the model LR-Cal CPB5600-DP is therefore a very pecisely manufatured piston-cylinder-system, which is loaded with masses in order to generate the individual test points.
The masses applied are proportional to the target pressure and this is achieved through optimally graduated weights. As standard, these masses are manufactured to the standard gravity (9.80665 m/s²),