Easily qualify Computed Radiography systems in accordance with the latest standards (EN 14784-1, ASTM E 2445).
Ultra-compact
With only 20 cm height and 27.5 cm width the smallest CR Phantom on the market
Ready to use
Including report template and test procedure
Well protected
A lined wooden case protects the CR Phantom during storage and transport
CR Phantom can test all relevant parameters of CR Scanner systems including basic spatial resolution, unsharpness, contrast, MTF, laser beam jitter, scanner slipping and shading. These tests demanded and described in detail in standards ASTM E 2445 and EN 14784-1 have to be performed periodically.
The DÜRR NDT CR Phantom exceeds these standards by including two Duplex wire type IQIs. Measuring points for shading correction are arranged in both axis directions - panorama and landscape. All required information is mapped on the imaging plate with a single X-ray exposure - the CR Phantom need not be rotated to generate the information of the second axis. This results in more accurate test scores and significant time savings.
• T-target - Brass
Laser Beam Jitter, MTF Check, Blooming (Flare)
• Duplex Wire Type IQI 15D - HiRes
Basic Spatial Resolution, Unsharpness
• BAM snail
Central Beam Alignment
• Line Pair IQI
Line Pair Resolution
• Measuring Points
Shading Correction
• Cassette Positioning Locator
Positioning of Cassette
• Homogeneous AL Strip
Scanner Slippage, Shading
• Lucite Plate
Carrier Plate
• cm/inch Ruler
Linearity Check
• Contrast Sensitivity Gauge
Contrast Sensitivity