The DigiMag® Series DM-1000 Digital Differential Pressure and Flow Gages monitor the pressure of air and compatible gases just as its famous analog predecessor the Magnehelic® Differential Pressure Gage. All models are factory calibrated to specific ranges. The 4-digit LCD can display readings in common English and metric units so conversions are not necessary. The simplified four button operation reduces set up time and simplifies calibration with its digital push button zero and span.
NIST CALIBRATION available in Product Configurator.
Product Applications
Filter monitoring
Air velocity or flow
Blower vacuum monitoring
Fan pressure indication
Duct, room or building pressures
Clean room positive pressure indication
Features
Field programmed reduces installation time
User selectable parameters for pressure, air velocity or flow permits same device for multiple applications
Specialized filter set point for alerts when maintenance is due
Security levels permit matches the correct access to right skill
Power versatility works with 9-24 VDC or 9 V battery allows deployment in a variety of spaces wired or not
Service:
Air and non-combustible, compatible gases.
Housing Materials:
Glass filled plastic.
Display Update:
Selectable for 1 second, 10 minutes or update only from button push.
Pressure Limits:
Normal and bi-directional ranges 5 in w.c. and lower = 2 psi (13.7 kPa); Normal and bi-directional ranges 10 in w.c. and higher = 11 psi (75 kPa).
Selectable Engineering Units:
in w.c., psi, kPa, Pa, mm w.c., mBar, in Hg, mm Hg, FS (0-100% FS).
Power Requirements:
9V battery, included, user replaceable or external power