The Autoclave Steam Sterilizer has a touch panel with Human Machine Interface (HMI), signal conditioning units, safety circuits and emergency switch. The Autoclave Steam Sterilizer is controlled at all time by the HMI. Which contains the operation automatic screen, manual, trend graph, parameter, dynamic and password, etc.
In order to destroy the uniqueness of microbes and bacteria, the Autoclave Steam Sterilizer is designed to withstand high-pressure steam and temperature, and use pure steam as the heating medium.
When the sterilization temperature is 121℃, the acceptable range of no-load and full-load's temperature and heat distribution is within plus or minus 0.5℃.
Features
The panel use LCD touch screen control.
Program Flow: PLC control. Easy operation. Providing three sterilization programs or customize specifications as customers' requirements.
Simple & Safe operation: The automatic program provides complete and safe control to ensure safe operating condition.
Temperature recorder for continuous monitoring and recording the sterilization process.
Vacuum filter: online SIP sterilization, completed test report, heating and drying for the filter shell to prevent the filter from clogging and moist bacteria.
The pure steam inlet can be used for condensation sampling, online inspection, dryness or steam quality test.
The inner carcass and interlayer pressure can reach 3.5 bar.
The heat distribution of inner carcass can reach temperature within ± 0.5°C.
Check the temperature of the holes, all the holes can drain water completely.
The inner carcass is a special design so that all water can be completely drained.