CL-201 Colorimetric Residual Chlorine Analyzer
Measurement principle:
Using DPD colorimetric method to detect the concentration of residual chlorine which is an measurement accurate, cost-effective and low maintenance instrument. It is suitable for the measurement of residual chlorine in chlorine disinfection process and monitoring of residual chlorine concentration in drinking water pipe network. Continuous operation 30 days of reagent dosage, realize unattended operation, operation and maintenance cost is low.
Features:
①The DPD standard colorimetric method
②The continuous operation of 30 days do not need to replace the reagent, simple operation, low operation and maintenance cost
③Two programmable relay, can set the concentration alarm
④It is suitable for measuring residual chlorine in tap water, circulating cooling water, sewage chlorination process of residual chlorine measurement and tap water pipe network.
Technical Indexes
Measure range
0~5ppm
Accuracy
±5% reading or ±0.03ppm Cl2,take big
Detection limit
0.03mg/L
Display
70mm×48 mm LCD
Calibration
Use default calibration curves
Operation condition
Sample flow rate: 200-400mL/min
Sample inlet pressure: 0.1~5bar
Sample temperature:5~40℃
Power supply
220-230VAC,50/60Hz
Output
4-20mA
Relay
Two groups of relays, can set the concentration alarm
Storage temperature range
-40~60℃
Storage humidity
humidity:10-90%, non condensing
Size
460mm*350mm*186mm(L*W*H)
Weight
About 9KG
Installation
Wall hanging