DOG-208 F Dissolved Oxygen Electrode applicable for Polarography Principle With platinum (Pt) as cathode and Ag / AgCl as anode The electrolyte is 0.1 M potassium chloride (KCI).The silicone rubber permeable membrane imported from U.S. serves as the permeable membrane.It has silicone rubber and steel gauze.It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape retention and other performances
DOG-13F Industrial Dissolved Oxygen Sensor
Features
DOG-13F Dissolved Oxygen Electrode applicable for Polarography Principle
With platinum (Pt) as cathode and Ag / AgCl as anode
The electrolyte is 0.1 M potassium chloride (KCI).
The silicone rubber permeable membrane imported from U.S. serves as the permeable
membrane.
It has silicone rubber and steel gauze.
It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape
retention and other performances
Technical Indexes
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: 60 days
Measured water temperature:0~60℃