Ellab Temperature Standard
ETS10
The traceable ETS allows for extremely accurate temperature measurements by converting resistance into temperature using its built-in algorithms.
When connected to the ValSuite software, the LiquiCal bath and ETS provide a perfect setup which is automatic. This saves time and allows for a convenient calibration.
Key Features and Benefits
Temperature Range -80 to +250 °C
Accuracy: ±0.015 °C / ±0.025 °C
Separated electronics from heat zone
Robust and Versatile Design
ETS10
Temperature Range: -80 to +250 °C
Accuracy – Including Instrument, PRT and Drift: -50 to +150 °C: ±0.015 °C
Other temperatures: ±0.025 °C
Measuring Sample Rate: 4 Sec.
Digital Filter: 0-10 Sec.
Resistance: 100 Ω
Sensor Response Time: T-63%: 33 Seconds
T-90%: 45 Seconds
Overall Dimension: ø 16 x 475 mm
Interface & Power Source: USB
Sensor Material: ANSI 316L
Dimension of Sensor: ø 6.35 x 305 mm
Minimum Immersion Depth: 100 mm
Environmental: 15 – 35 °C / 0 – 90% RH, non-condensing
Calibration: Traceable to NPL
Calibration Algoritms: ITS-90
Recommended Calibration Interval: 12 months
Weight w / wo Display: 0.35 kg / 0.25 kg
Optional Display
Display: 8 digits, 7 segments, green LED
Resolution: 0.001°C