The MIPO 5 SERVOCOULOMETER, is an analyzer used to measure total chlorine (and other halogens) by means of coulometric method.
Its measuring capacity extends from traces ( 80%). It is also used to measure chlorine On Chlorinated PVC-C. In addition to the accessories for measuring EOX and POX, we offer complete equipment for AOX dosing by agitation and absorption.
Chlorine and AOX analyzer according to the standard ASTM D4929, ASTM D5194, ISO 9562...
This chlorine analyzer has two furnaces, and ensures complete combustion of your samples. The furnace F1 constitutes a real micro reactor, allowing a great flexibility of use. It allows you to program controlled pyrolysis and combustion of your samples through temperature sequences. With this technology, you will no longer generate soot regardless of the nature of the sample and the position of the furnace (vertical or horizontal).
Principle: Combustion - Coulomety
Detection Limit: < 0.1 ppm to more than 80%
Quantification: < 0.3 ppm to more than 80%
Accuracy / RSD:
At 0.4 ppm: 7.51 %
At 11 ppm: 0.79 %
At 1000 ppm: 0.46 %
Injections:
20 to 80 µl liquids
10 µL LPG
0.5 to 100 mg solids