Mass flow meter ST75 series
for fuelfor gasfor air

Mass flow meter - ST75 series - FCI Fluid Components International - for fuel / for gas / for air
Mass flow meter - ST75 series - FCI Fluid Components International - for fuel / for gas / for air
Mass flow meter - ST75 series - FCI Fluid Components International - for fuel / for gas / for air - image - 2
Add to favorites
Compare this product
 

Characteristics

Type
mass
Fluid
for fuel, for gas, for air
Installation
in-line
Material
stainless steel, aluminum
Communication
4-20 mA, Modbus, HART
Protection level
rugged, ATEX, IP67, IP66, SIL
Applications
industrial
Other characteristics
compact, precision, remote
Volumetric flow rate

Max.: 951.57 m³/h
(33,604.3774 ft³/h)

Min.: 5.13 m³/h
(181.1642 ft³/h)

Process temperature

Max.: 121 °C
(250 °F)

Min.: -18 °C
(-0 °F)

Process pressure

16.5 bar, 41 bar
(239.31 psi, 594.65 psi)

Precision

0.5 %, 1 %, 2 %

Repeatibility

0.5 %

Description

The ST75 Series thermal flow meters are a compact, low cost solution for accurate, repeatable direct mass flow measurement of air and gases. These in-line (spool-piece) style flow meters have no moving parts and are available for use in pipe diameters from 1/4 inches to 2 inches, [6 mm to 51 mm]. Process connection options include male NPT, female NPT, and flanged. The ST75 Series electronics are housed in rugged, IP67 rated enclosure with dual conduit ports in either NPT or M20 threading. The instrument comes standard with dual 4-20 mA outputs and a 500 Hz pulse output. The models ST75A and ST75AV include HART or Modbus, as well as NAMUR compliant 4-20 mA outputs and SIL compliance rating. The flow meter's transmitter/electronics can be integrally mounted with the flow element (probe) or be remote mounted to best match your installation situation. The complete instrument carries global Ex agency approvals for Division l/Zone 1 installations. The ST75 flow meter sensing element features precision, platinum RTD in small diameter, "equal-mass," all-metal thermowells to provide superior accuracy, repeatability and fast response. The "V" suffix models (ST75V and ST75AV) include Vortab® flow conditioners built-in to the spool-piece to provide accurate and repeatable flow measurement installations with limited straight-run and/or for operating in transitional flow ranges. Applications Burner and boiler fuel and air feed lines Industrial furnaces, kilns and oven fuel/air controls Natural gas sub-metering Chiller air flow measurements Dosing and gas injection rate controls Heat treating gas controls industrial furnaces, kilns and oven fuel/air controls

Catalogs

No catalogs are available for this product.

See all of FCI Fluid Components International‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.