Single stream flow computer C581
for superheated steamsaturated steamfor coolant

Single stream flow computer - C581 - Fluidwell bv - for superheated steam / saturated steam / for coolant
Single stream flow computer - C581 - Fluidwell bv - for superheated steam / saturated steam / for coolant
Add to favorites
Compare this product

Characteristics

Applications
for superheated steam, saturated steam, for coolant, for water
Type
mass flow, volume, flow power
Options
temperature-compensated, data logger, differential, with pressure compensator, single stream

Description

The C581 flow and energy computer measures the actual flow, temperature and pressure and uses the IAPWS determinations or Water-Glycol tables to calculate the energy, mass and volumetric flow. The linearization enhances accuracy in any flow range. The large logging register contains all measured and calculated values. The robust IP66/67, Type4X aluminum field enclosure, withstands extreme weather and industrial conditions. Our proven experience with instrumentation, international requirements and the focus on user-friendly operation are combined in the most robust, accurate and simple field mount Flow & Energy Computer on the market. Advantages - Large, versatile display with bright backlight, all info at a glance. - Easy-to-operate with 6 rugged industrial silicone buttons. - User-friendly operation and configuration via the keypad and display. - 2 temperature inputs for inlet and outlet measurement, e.g. at a heat exchanger. Features - Displays calculated mass flow, volumetric flow and power at actual conditions. - Calculates totalized values for volume, mass and energy. - Determines density and enthalpy according to IAPWS. - Main screen with custom selectable order of measured and calculated variables. - Data logging of process parameters and calculated values. - 11 linearization points per input, for all inputs. - Ability to process various flow inputs: pulse (pending), (stacked) differential pressure or other 4 - 20mA signals. Applications - For industrial environments, measuring coolants, water, superheated and saturated steam at heat exchangers, boilers, condensers and coolers. - For single or stacked differential pressure based flowmeters or averaging systems.

VIDEO

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.