• IEC62133-2017
Performance parameters:
Inner Chamber Dimension: W600 X H750 X D500 MM
Inner Chamber Material: Stainless steel with mirror-finish, bottom reinforcement
External Chamber Material: Cold-Rolled Steel with paint-baked coating
Observation Window: The single-opening door is accompanied by a 3-layer vacuum large glass observation window (W430mm x H 580mm); The door frame is equipped with electric heating (automatic adjustment) anti-frost, anti-condensation device; so that you can clearly observe the Test status of the test sample in the box through the observation window at any time.
Internal Short Circuit Technical Parameters: Max. Loading200KG
Resolution: 1/10000
Force Accuracy: ≤0.5%
Displacement resolution: 1/1000
Accuracy of displacement: ≤0.5%
Speed Range: 0.01-5MM/S
Effective Range: 0 ~ 200mm
Max. Test Area: 300X300MM (Can be customized)
Voltage Range: 0—60V
Voltage Resolution: 5mV:
Voltage Acquisition Frequency: 100HZ
Control System: Servo motor + Servo control system
Safety Protection: Explosion-proof pressure relief device, smoke exhaust device, fire extinguishing system, electronic door lock system, safety stop switch