The P20 and P40 are optical incremental linear encoders or measuring gauges. The encoders are capable of measuring travel lengths of at most 40 mm and can be placed in a Quality Control Laboratory or on a factory floor. They have ball bearing plunger guides to smoothly operate and highly resist side loads. They also have an internal ASIC to interpolate electronically and reliably.
The Models P20 and P40 Measuring Gauges are plunger-type linear optical encoders. They areideal for precise measurement of travel lengths up to 40 mm, and are equally at home in the QClab or on the factory floor.The P20 and P40 incorporate ball-bearing plunger guides for smooth operation and greaterresistance to side loads, and an internal ASIC for reliable electronic interpolation. The plungeris spring-loaded to return to the fully extended position.
Measuring length:P20: 20 mm (0.79”); P40: 40 mm (1.57”)Resolution:10, 5, 2, 1 0r 0.5 micrometers* Accuracy at 201.0 m standard;Power required:+ 5 VDCMeasuring force:1.2 N max (5 oz)Allowable side load:4.4 N (1.0 lb)Mounting method:Through clearance holes against a flat surface, or with standard 8h6clamping shankMeasuring contact:Corundum with 10 mm (0.4") spherical tipProtection Case:64Plunger:IP50Operating temp.:0 C to 50 C (32 F to 122 F)Storage temperature:-20 C to 70 C (-4 F to 158 F)Humidity:10 to 90% rh, non-condensingCable length:1 m (40") standardff£IPooo oooo o£oC:5%, 120 mA maxMaximum speed:1 m/s (40 in/s)±±mm±±0.5 mData format:Quadrature square wavesIndex format:1/4-cycle wide, gated with high states A and BIndex location:5 1 mm