Quantaurus-QY® Plus (UV-NIR absolute PL quantum yield spectrometer C13534-11,-12) is designed to measure photoluminescence quantum yield using the absolute method, instantaneously.
Two types have been made available, C13534-11 (300 nm to 950 nm) and C13534-12 (400 nm to 1100 nm), according to the measurement wavelength region. Additionally, easily extended functions using newly designed options allows measurements to be made in the near-infrared region up to 1650 nm, at high sensitivity (evaluation of low quantum yield), and for upconversion emissions.
C13534-11 is designed to measure the instantaneous absolute value of emission quantum yield using the photoluminescence method. (300 nm to 950 nm) . Additionally, easily extended functions using newly designed options allows measurements to be made in the near-infrared up to 1650 nm, at high sensitivity (evaluation of low quantum yield), and of upconverted emissions.
Specifications
• PL measurement wavelength range : 300 nm to 950 nm
• Monochromatic light source:Light source : 150 W xenon light source
• Excitation wavelength : 250 nm to 850 nm
• Bandwidth : 10 nm or less (FWHM)
• Excitation wavelength control : Automatic control
• Multichannel spectroscope:Photo-detector : BT-CCD linear image sensor
• Measurement wavelength range : 200 nm to 950 nm
• Wavelength resolution : < 2 nm
• Number of photosensitive device channels : 1024 ch
• Device cooling temperature : -15 ˚C
• AD resolution : 16 bit
• Spectroscope optical arrangement : Czerny-Turner type
• Integrating sphere:Material : Spectralon
• Size : 3.3 inch