SM86 sterility test system is used for sterility test in sterile product.It can work with all kinds of closed canister. Many types of sterile products can be tested with the SM86 sterility test system, including:
◆ LVP(Large Volume Parenteral) in Glass bottle, plastic bottle, soft pouch, etc.
◆ Injection in Ampoule and vial, and freeze-dried powder
◆ Insoluble powder
◆ Medical devices
◆ Preloaded syringe
It is to test whether testing sample contains microorganisms. The microorganisms present in the sample are captured on the microporous membrane in the closed canister by the action of directional peristaltic pressure from the SM86 pump device. Rinsing membrane to remove the bacteriostatic ingredient if samples included. Appropriate medium is pumped into each canister separately to promote the growth of the captured organisms. The canisters are incubated and examined for contamination in accordance with the relevant pharmacopoeia.