Product Description
Immersion UV is not suitable for the occasion of poor water quality conditions, such as sewage treatment plant entry),In these situations, the optical window surface is contaminated in a few minutes, every time the extraction UV water quality analyzer measures through extrating liquid into the measuring chamber by the peristaltic pump. After the measurement, the water can be discharged, and measuring pool should be cleaned by clean fluid.
Features
Measuring pool immerses in the cleaning fluid at ordinary times, real-time sampling, real-time emptying, highadaptation to thepoor water quality
Automatic zero, no maintenance, no scale calibration, only regular cleaning fluid
Full spectrum (200-800nm) measurement techniques and chemometric algorithms can measure more parameters
an realize large range measurement (COD up to 10000 mg/L)
Fast response(5s) is applicable to real-time requirements of the occasion.
Interference-free chlorine ion
Parameters
Method DOAS
Range (COD) The lowest: 20mg / L (optical path 16mm),
The maximum: 10000mg / L (optical path 4mm)
Zero Drift <1% F.S.
Linearity ± 2% F.S.
Measuring Cycle the shortest 60 seconds, the cycle can be set
Water Temperature Requirement 0 ℃~45 ℃
External Interface measuring module, RS232,The whole machine,RS232/485, 2×4-20 mA output
Measuring Module Size 176 mm×144 mm×96 mm
The Whole Nachine Size 283 mm×217.5 mm×368 mm
Measurement Module Power 12 VDC, 10W
The Machine Power Supply 220VAC, 100W
Applications
Apply to the occasion of poor water quality conditions, for process on-line detection such as sewage enterprises export, sewage treatment plants, sewage treatment plant entrance