Configuration
The PMLA electronic multi-channel load has up to 3 cooling units, each with 4 mounting positions for load modules. 150 W, 300 W, 450 W or 600 W load modules are available. Depending on power, a module occupies one (150 W), two (300 W), three (450 W) or four slots (600 W).
Load Modules
The modules are available in four different voltages 40 V, 60 V, 120 V and 240 V and currents from 1 A up to 120 A. Various loads can be configured, e.g.: 1 x 600 W + 1 x 450 W + 2 x 300 W + 5 x 150 W The total power is max. 1,800 W. The loads can therefore be easily configured to test units with multiple outputs. The load inputs are galvanically separated. Very simple systems can therefore be specially configured for requirements with multi-channel Burn-In equipment.
Operating Modes and Functions
The following operating modes are possible:
•Current mode
•Voltage mode
•Resistance mode
•Power mode
•Dynamic mode with up to 100 setting values
An adjustable current or voltage protection as well as voltage and current measurement functions are available.
Measurement Data Storage
A data acquisition function records up to 100 measurement points (time, voltage, current) at each channel and saves it internally.
User Interface
The master device has a graphical user interface. All channels, also those of connected slave devices, can be operated.
I/O Port
I/O Port for the following functions.
•Load setting C and V
•Load on/off
•Voltage monitor output
•Current monitor output
•Sense inputs