Introduction
Gasboard-5230 Automobile Emission Gas Analyzer adopts micro flow NDIR principle with independent intellectual property and international PCT (CN2018100767) to detect NO, dual channel structure sensor and special filters highly resolve environmental temp. instability and electroniccomponents aging issues. It adopts NDIR principle to measure CO2 and NDUV principle to measure NO2 directly without NOx converter.
Features
• Micro-flow NDIR NO gas bench replaces traditional ECD NO sensor, high accuracy, longer lifetime, no regular replacement.
• NDUV NO2 gas bench without NOx converter, direct NO2 detection with high accuracy.
• Auto temp. and pressure compensation
• Multi-stage filtration system ensures longer lifetime
Specifications
Gas Measurements: CO2 NO NO2
Range: CO2: (0.00-20.00) %; NO: (0~5000) ppm;NO2: (0~1000) ppm
Accuracy:
--CO2: (0.00~16.00) %:Abs. error:±0.3%;Rel. error:±3%; (16.01~20.00) %: Rel. error:±5%
--NO: (0~4000) ppm:Abs. error:±25 ppm;Rel. error:±4%; (4001~5000) ppm:Rel. error:±8%
--NO2: (0~1000) ppm:Abs. error:±25 ppm;Rel. error:±4%;
*Remark: Abs. error or Rel. error whichever is greater
Resolution: NO/NO2: 1ppm;CO2: 0.01%;
Response Time: HC/CO/CO2: T90≤8s;NO/NO2: T90≤10s;O2: T90≤12s
Warm up Time: 30min
Communication: RS-485/232