Battery monitoring unit P/N 500-100-106
chemicalairalarm

Battery monitoring unit - P/N 500-100-106 - HÜRNER Schweisstechnik - chemical / air / alarm
Battery monitoring unit - P/N 500-100-106 - HÜRNER Schweisstechnik - chemical / air / alarm
Add to favorites
Compare this product
 

Characteristics

Applications
battery, chemical, air, alarm

Description

This monitoring unit provides easy airflow monitoring of chemical storage cabinets. A sensor monitors the extraction airflow and triggers a visual and audible alarm when no airflow above the defined minimum is detected. The air monitoring unit without fan ships preassembled to be installed immediately on top of the storage cabinet. The alarm threshold for the airflow velocity is factory-set, but can be customized at the differential pressure meter/switch. The enclosure is made of silver pearl glimmer, powder-coated sheet steel. The airflow sensor monitors the actual extraction velocity and the entire extraction system of the storage cabinet. Both the volumetric flow and the power supply are sensed by a microcontroller unit. When airflow velocity drops below the set threshold, visual and audible alarms are set off. The system ships with two potential-free (“dry”) contacts (24  V, 2 A) that can be used to send an alarm signal to the control center. An integrated battery cuts in in case of power supply failure, and the monitoring system and the display, including simple error messaging, can run on it for 4 hours. The product ships ready for plug-in. DIMENSIONS (W X H X D) - 210 x 215 x 210 mm RUBBER SLEEVE - 75 mm POWER SUPPLY - 230 V, 50 Hz CURRENT INPUT - 0.1 A

Catalogs

No catalogs are available for this product.

See all of HÜRNER Schweisstechnik‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.