Application
SDTGA8000 can be used to determine the moisture, ash, volatile matter, LOI (loss on ignition) as well as calculation of fixed carbon and calorific value in coal, coke, solid wastes and other solid biomass fuel.
It can also be used for the calculation of carbon content in fly ash and clinker.
Specification
Test index Moisture, ash, volatile matter and LOI
Standards ISO11722, ISO1171, ISO562, ASTM-D3173 / D3174 / D3175 / D5142 / D7582
Precision 0.02% RSD
Furnace Dual furnaces, one for moisture, one for ash and volatile matter
Sample weight 0.5 up to 5g
Furnace temperature Room temperature~1000℃
Sample Loader 18 samples
Temperature control precision ±2℃ (Moisture), ±5℃ (Ash, Volatile matter)
Test efficiency Full indexes for 3 batches of samples (54 samples) within 8h
Power requiremen 220 V(-15%~10%),50 / 60Hz
Gas requirement Compressed air / Oxygen & Nitrogen
Max.Power 4.5kW
Highlights:
High precision & accuracy.
①Two internal balances away from high temperature and heat radiation, no drift occurs.
②With automatic calibration function, each weighing is accurate and reliable.
③Simulate test environment and condition of drying oven and muffle furnace, exact in accordance with standard.
④The unique automatic cleaning function of low temperature furnace completely eliminates the influence of dust on result.
High efficiency
①Dual furnaces design, low-temperature furnace for moisture test and high- temperature furnace for ash and volatile matter test.
②Analysis for moisture, ash and volatile matter can be conducted simultaneously and consecutively.