Applications:SDTGA6000 can be used to determine the moisture, ash and volatile matter as well as calculation of fixed carbon and calorific value in coal, coke, calcined petrol coke and other solid biomass fuel etc.
Specification
Test indexMoisture, ash, volatile matter and LOI
Standards ISO11722, ISO1171, ISO562, ASTM-D3173 / D3174 / D3175 / D5142 / D7582
Precision 0.02% RSD
Furnace Dual furnaces, one for moisture and ash, one for volatile matter
Furnace temperature Room temperature~1000℃
Temperature control precision ±2℃
Temperature resolution 0.1℃
Balance Resolution 0.0001g
Max. Samples 19 samples + 1 blank sample position (Moisture & Ash)
25 samples + 1 blank sample position (Volatile matter)
Quantity of sample / test time Full indexes for 19 samples< 120 min (Quick method)
Sample weight 0.5 up to 5g
Max. Power 4kW for moisture/ash testing; 2.5kW for volatile matter testing
Highlights:①Dual furnaces structure, the system consists of two parts, one for moisture and ash analysis, other for volatile matter analysis.②For moisture and ash analysis, 19 samples can be loaded once + 1 blank sample position;③For volatile matter analysis, vertical furnace structure is adopted with 26 positions. Upper part for high temperature volatile matter test while lower part for low temperature cooling and weighing, makes the test result more accurate and test efficiency greatly improved.④Flexible software design, test procedure, analysis time, temperature etc. can be customized;