Charge mode, high impedance piezoelectric measurementdemands highly insulated coaxial cables and connectors toensure an insulation resistance greater than 1013 Ω throughoutthe measuring chain. Only low noise coaxial cables that produce very little triboelectricity during movement may be used.The connectors must be robust, sealed and resistant to dirt.Kistler connectors have been developed specifically to meetthese requirements and are made of stainless steel. Unlike galvanized connectors they are therefore not subject to any wear,and measurement reliability and accuracy are improved. AllKistler connectors contain an O-ring seal at the cable end andthe connection end.
• Connectors: KIAG 10-32 pos. (sensor), BNC pos. (amplifier)
Certificates and Standards
Degree of protection EN 60529
IP65, IP40
Cable properties
Interfaces
Connector type
10-32 UNF (positive), BNC (positive)