All-rounders for general use
Ideal for working with organic solvents and cold or diluted acids. Depending on the ceiling height of your laboratory, you can choose between tall or low designs.
Fume cupboard for low ceiling height with unique low installation height (2250 mm)
Above-average usable worktop depth
Underbench storage for tall utensils up to 720 mm
Metallic construction enables columns with hollow spaces at the front – for easy maintenance and protection against air leakage
Interior: coated steel, optional lining in Trespa®
Side or front service connections for easy tapping of water, gas, etc.
Width
1200
Front height : 2300/2550
Chamber height : 2300/2600
Depth : 900
Usable space width : 1100
Usable space height : 1190/1490
Usable space depth : 800
Work height : 900
Minimal permissable volume flow rate* : 450
Specified volume flow rate : 480
Maximum permissible volume flow rate : 1500
our work is what matters
When selecting a fume cupboard, the most important considerations are the activities you do and the substances you use.
The certification according to DIN EN 14175 part 2 and 3 allows EXPLORIS® fume cupboards for general use to be used universally in the laboratory. Below you will find all information about dimensions and volume flow.
Fume cupboard control TouchTronic
Easy operation of fume cupboards
As a control element for a wide range of functions, the new TouchTronic combines fume cupboard control and monitoring with high computing power. The combination of work steps in one unit saves time and also protects everyone else in the lab: warnings can be easily shown on the display.