•Nodes and parts being in contact with steam, distilled and double distilled water are made of AISI 321 stainless steel and other materials which do not affect the quality of product water.
•Low specific power consumption – 1.51 kW for 1 l of double distilled water
•Low level of cooling water consumption - 24.2 liters per 1 liter of double-distilled water.
•Built-in control unit.
•Source water supply tube and distillate drain tube are included into the kit.
•Reagents can be added to the evaporation chamber at the 2nd distillation stage for double distillation at their presence. The reagents shall be inert to stainless steel and PVC plastic.
•Protective panels protect personnel from contact with hot parts and have decorative purpose.
•Heating elements switch off automatically if centralized water supply disrupts and water level in the evaporation chamber drops below the allowable level.
•Level of water in the distiller fed for evaporation is maintained automatically.
Assignment double distilled water production
Capacity, l/h distillate: 13.8; - double distillate: 12.0
Electrical conductivity of product water double distillate: 1.0-1.2 μS/cm; - distillate: 5.0 μS/cm max
Quality of product water Type II, III
Product water temperature, °С- distillate: from 70 to 85; - double distillate: 40 max
Current type, frequency, voltage three-phase AC, 50 Hz, 380 V
Power consumption for obtaining of: - distillate: 8.1 kVA; - double distillate: 15.3 kVA
Source water consumption, l/h (at water t from 10 to 12 °С at pressure from 0.2 to 0.4 MPa) for obtaining of: - distillate: 104.0; - double distillate: 240.0
Source water pressure, MPa 0.1 – 0.4