Equipment for calibration of MAV testers
With the use of the MAV calibration equipment you are able to calibrate and adjust MAV devices on your own. This way, downtimes are reduced and costs minimized long-term wise.
For a successful calibration / adjustment of your test stations you need the following equipment:
KMG-force gauge as a reference system.
Load cell(s) “SM” for the respective force range.
Fixtures for the load cell(s) for the assembly on the test station.
Optional: mavCALIB-software as a support and for the creation of certificates.
The calibration process is as follows:
Using the fixtures, the load cell has to be assembled on the test station that is subject to calibration. Load cell and KMG have to be connected using the data cable. Subsequently force is applied to the machine in predefined steps and displayed values of the machine and the KMG reference system, which receives the correct force values via the load cell, have to be compared (calibration). The internal parameters of the machine should then be adjusted within the setup-menu until the displayed values of machine and KMG match (adjustment). A detailed manual for the calibration of all models can be found within the KMG’s operation manual. For information regarding mavCALIB and a downloadable trial version navigate to “Products” - “Software”.
KMG-force gauge and SM-load cells
Up to eight different load cells with different force ranges may be run via one KMG device.
Very high resolution of 10,000 steps (equals 0.005 N in case of 500 N load cell).
Compact and easy-to-handle construction.
MKG with modern TFT-touchscreen.