Model M151 is a stable high current calibrator up to 120 A. Basic accuracy is 275 ppm. Calibrator can work in a simulated amplifier mode to increase current ranges of any multifunction calibrator. It is suitable for power meter’s calibration because M151 can be synchronized with the input signal not only in amplitude but also in frequency and phase. Current terminals are isolated up to 450 Vpk against case (protective earth). M151 is a sophisticated instrument with its own recalibration procedure. The procedure enables to adjust any deviation directly from the front panel. Calibrator is designed for checking parameters of amp meters. With current coil it can be used for calibration of clamp meters.
RS-232 and GPIB interface allow for remote control of the calibrator and time saving automated calibrations using SW package CALIBER/WinQbase.
Specifications:
DC current: 8 mA - 120 A (± 250 ppm / year)
3000 A with 151-25 Current Coil
AC current: 8 mA - 120 A, 15 Hz - 1 kHz (± 250 ppm / year)
3000 A with 151-25 Current Coil
Process multimeter:
DC voltage: 0 - 20 V (± 100 ppm / year)
AC voltage: 1 - 20 V (± 200 ppm / year)
DC current: 0 - 200 mA (± 100 ppm / year)
AC current: 10 - 200 mA (± 200 ppm / year)
Frequency: 1 Hz - 10 kHz (± 50 ppm / year)