Incubator Im IN75mplus
The incubator Im is a Class I medical device.
Standard delivery Incubator Im
The heating of this incubator is optimally tuned for natural convection and valuable chamber loads for research, pharmaceutics, medicine and food chemistry are warmed up very carefully.
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable temperature values
adjustable parameters: temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Sterilisation: fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load
Ventilation
Convection: natural convection
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
Vent: vent connection with restrictor flap
Communication
Documentation: programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port