FlowPrep 060 lets you choose the temperature, gas, and flow rate best suited for your sample material and application.
The FlowPrep™ 060 applies both heat and a stream of inert gas to the sample. The heat causes contaminants to desorb from the surface and the stream of inert gas sweeps them out of the sample tube. It lets you choose the temperature, gas, and flow rate best suited for your sample material and application. Needle valves allow you to introduce the flowing gas slowly to prevent fluidization of samples.
Specifications
Sample Tubes: Up to 3.0-cm (1.18-in.) OD
Heating Block:
Temperature: 10 ºC above ambient to 400 ºC
Temperature Accuracy: ± 10 ºC
Electrical
Voltage: 100, 120, 220, 240 VAC ± 10%
Power: 250 VA maximum
Frequency: 50/60 Hz
Environment
Temperature: 10 to 35 ºC operating; 0 to50 ºC non-operating
Humidity: 45 to 80% relative, non-condensing
Gases Used: Helium, nitrogen, argon, or any other pure gas that does not adsorb at room temperature
Flow Rate: Up to 50 cm3/min. No vacuum required
Sample Capacity: 6 heating stations and 6 cooling stations
Physical
Height: 45 cm (16.93 in.)
Width: 36 cm (14.17 in.)
Depth: 30 cm (11.81 in.)
Weight: 8.6 kg (19 lbs)