Applications:
Mechanical equipments and experimental facilities.
Technical data:
Connecting cable:Silicone, temperature range:-50℃~+180℃
Teflon, temperature range -50℃~+260℃
Glass Fiber, temperature range -50℃~+350℃
Protection tube: Stainless steel 304/316
Sensor element: Pt100, Pt500, Pt1000 meets IEC 751
Tolerance : Class B:△t=±(0.30 + 0.005 |t| )℃(standard)
Class A:△t=±(0.15 + 0.002 |t| )℃
Response time: t
0.9 approx. 10secs, in water 0.2 m/sec, diameter 6mm