This version of the High-Power Xenon Light Source Installation and Operation Manual corresponds to the units manufactured in the United States. They are identifiable by serial numbers beginning with serial number 10500 for the HPX-2000 and serial number 20500 for the HPX-2000.
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to your system. To facilitate these changes, you must first contact Customer Support and obtain a Return Merchandise Authorization (RMA) number. Please contact Ocean Optics for specific instructions when returning a product.
Overview
The following sections provide instructions on unpacking and setting up your HPX-2000 HighPower Light Source. Before using the HPX-2000 for the first time, check for transport damage. Be sure to adhere to all warnings on the unit and in this manual.
► Procedure
1. Unpack your lamp assembly carefully. Although the lamp is rigidly mounted, dropping this instrument can cause permanent damage.
2. Inspect the outside of the instrument and make sure that there is no damage. Do not use the instrument if damage is present. Contact your dealer for repair or replacement information, if necessary
Contents Your HPX-2000 package should contain the following:
HPX-2000 High-Power Light Source
Power cord
UV safety goggles
One IC-DB15-2 interface cable for shutter operation
Quick Start
The following table provides information on optimizing the operating environment of your HPX2000.
Moisture 5-95% without condensation at 40°C
Ventilation Unit should be situated so that its location or position does not interfere with proper ventilation.