Collision Gas Add-On
Designed to work exclusively with Genius XE 35 and Genius XE 70, our Collision Gas Add-On supplies Ultra High Purity Nitrogen to the collision cell at a flow of up to 200cc/min.
With a set purity of 99.999%, the Collision Gas Add-On produces a high quality flow with no reduction to the output of the Genius XE generator. The Collision Gas Add-On allows labs to draw nitrogen from their Genius XE 35 or Genius XE 70 which removes the need for gas cylinders for UHP nitrogen.
Features
• Can be used to supply Ultra High Purity Nitrogen for the collision cell - 99.999% nitrogen at 200cc/min*
• Designed to be used alongside a Genius XE 35 or XE 70, creating a dual purity system
• HMI Screen – monitor the performance of the Collision Gas Add-On
• Maintenance interval every year, down time minimized as service can be performed at same time as Genius XE
• No risk of running out of gas in the middle of analysis
• 2 year comprehensive manufacturer’s warranty
Site & Installation
• Gas Outlets Fitting: 1 x 1/4” BSPP
• Min Air Inlet Flow: 4 L/min
• Voltage: 100v - 240v
• Frequency: 50/60 Hz
• Current: 0.2A
• Max Operating Temp: 35°C / 95°F
• Particles: <0.01µm
• Phthalates: NONE
• Noise Level: Silent in operation
Dimensions & weight
• Size (HxWxD) mm: 165 x 360 x 425 mm
• Size (HxWxD) Inches: 6.5 x 14.2 x 16.7 inches
• Generator Weight: 14kg / 30.8lbs