The Infinity 1045 Compressorless Nitrogen Generator is specifically designed to supply carrier nitrogen gas and a separate dry air output to the Bruker EVOQ triple quadrupole mass spectrometer.
Features
• Works according to Bruker EVOQ triple quadrupole mass spectrometer specifications and is suitable for instruments with similar requirements
• Operate 24/7 with optimal performance when you need it
• Since the gas is supplied on demand, The generator will work according to your schedule.
• Few moving parts ensure long generator life with little maintenance required
• Minimal setup required Completely silent during operation
• 12 months comprehensive on-site warranty
Site & Installation
• Min Inlet Pressure: 116psi/8bar
• Max Inlet Pressure: 145psi/10bar
• Min Air Inlet Flow: 215
• Min Inlet Air Quality: 1.4.1
• Start Up Time: 60 mins
• Power Consumption: none
• Voltage: N/A
• Frequency: N/A
• Current: N/A
• Heat output: none
• Max Operating Temp: 30°C / 86°F
• Particles: <0.01 μm
• Phthalates: none
• Suspended Liquids: none
• Accreditations: CE
Dimensions & weight
• Size (HxWxD) mm: 730 x 424 x 250 mm
• Size (HxWxD) Inches: 28.7 x 16.7 x 9.8 inches
• Generator Weight: 39kg / 85.8lbs