video corpo

AC ammeter TA325
three-phase

AC ammeter
AC ammeter
Add to favorites
Compare this product
 

Characteristics

Technology
AC, three-phase

Description

The TA325 is a 3-phase AC current probe utilizing the Rogowski coil principle, used to measure AC current up to 3000 A when used with oscilloscopes, recorders or data loggers. The flexible probe allows current measurements on hard-to-reach conductors. Rogowski coil technology has a couple of advantages over CT (current transformer) technology: when measuring multiple phases, the phase difference is small and predictable for higher frequency harmonic analysis applications; also, Rogowski coils do not have magnetic materials to saturate and so have excellent linearity. The probes provide a 3 V full-scale AC output proportional to the current being measured with three selectable ranges: 30 A, 300 A and 3000 A. When using the TA325 with a PicoScope 4444 differential scope, ensure that you purchase three TA271 D9-BNC adaptors. These can be found under the Accessories tab on this page.
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.