Product Usage:
Ozone aging resistance equipment for testing rubber and its products (satisfying) relevant standards: JIS K 6259 ASTM1149 ISO1431 GB/T7762 GB/T13642, etc... mainly used as a toolbox for antioxidant testing;
2.Performance introduction:
Ozone concentration: 20~1000 pphm (set direct reading) Ozone concentration deviation: GB 10% Test chamber temperature: Measurement range Room temperature~+80°C Temperature uniformity: ±2°C Relative humidity: Measuring range 60% or more Speed: 0~50mm/s Dynamic pull-up frequency: 0.5Hz New sample rack (each fixture can be installed with tensile test piece to meet the type I dumbbell piece and actual workpiece) Test gas flow rate: 0-80L/min.
3.The test methods are divided into dynamic and static tests:
1、Speed-regulating motor drives the sample holder device (satisfying the static tensile test, the running track of the test piece satisfies the principle of the cycloidal path)
2、Ozone concentration automatic digital display controller.
3、Dual beam "QUA" non-dispersive UV ozone detector.
4、Digital display automatic online measurement.
5、 There is a metrology verification correction interface.
6、The ozone concentration/temperature and humidity setting
value controls the PID self-tuning, and the operation is
convenient and reliable.