Metal high temperature tensile test
Metal low temperature tensile test
Rubber high & low temperature tensile test
Other material high & low temperature tensile test
Test standard:
GB/T4338-1995 ISO783:1989 metal high tempertature tensile test
GB/T13239-2006 ISO15579:2000 metal low temperature tensile test
GB/T228-2002 ISO6892:1998 metal room temperature tensile test
Specification:
Capacity - 500N, 1KN, 2KN, 5KN, 10KN, 20KN, 50KN can option
Resolution - 1/500.000
Load accuracy - ±0.25%F.S
Test speed range - 0.001-500mm/min or 1-1000mm/min
Temperature range - Room Temperature~250℃ - Room Temperature~-40℃ - Room Temperature~-70℃ - -40℃~150℃ - -70~200℃
Inner chamber dimension - W300xD350xH600mm
Temperature control accuracy - ±2℃
Coolants - Low temperature chamber, environmental coolants
Chamber material - Inner side will be SUS304# mirror, the outside will be cold steel plate with painting
Temperature control mode - P.I.D self-adjusting mode and SSR control
Temperature controller - Button input, LED digital display output
View window - Double layer view window(180x220mm)
High-precision
Use Panasonic AC servo motor to drive, and ABBA high precision ball screw for transmission, with high precision explosion proof LOADCELL from U.S. TRANSCELL, the force accuracy can be ±0.25%, the displacement accuracy is up to 0.001mm.
High intelligence
Automatically determine the fracture,automatic return, the force units are free to switch languages switchable:Simplified/Traditional Chinese, English. Load-displacement, load-time, displacement-time stress-strain, strain-time, stress-time multi-curve analysis.