Flow-through tubes with funnel and riser are available with integrated:
Temperature sensor
Thermostats and additional hose connection for water circulator
Tube for manual sample changing, not thermostable
Only for Unipol, Polartronic V and H series, Saccharomat models with plug-in connection inside of the sample compartment
Polarimeter tubes of SCHMIDT + HAENSCH are designed for different applications. All tubes meet the ICUMSA recommendations, class A (±0.01%) and the corresponding specifications of the O.I.M.L. and the Australian Standard K 157 regarding construction and precision. In general: Longer measuring tubes are more reliable than short ones. Sample heterogeneity and residues from previous samples have a larger influence on the accuracy of the measurement when applying shorter tubes. Larger sample volumes ensure more stability. An exception is the case of dark samples; a shorter measuring tube might be necessary due to the high absorption of light.
Checkpoints
Simple filling with funnel
Available with thermostat hose connection
With integrated temperature sensor
Ideal to measure large sample quantities in flow-through