Test equipment «Dynamics» is to test gas, vapor, and combined filters of half and overall masks according to EN 405, EN 1827, EN 12941, EN 12942, EN 14387 using test agents:
1) Gases (chlorine, sulfur dioxide, ammonia, dimethyl ether, isobutene, hydrogen sulfide, nitric oxide, hydrogen chloride, arsenic hydrogen)
3) Vapors (cyclohexane, hydrogen cyanide, acrolein, benzene)
4) Hydrargyrum
Operation:
Test equipment prepares the gas-vapor-air mixture in accordance with the conditions for using the filter: the required temperature, relative humidity, and concentration of the test agents.
In the test chamber, the mixture passes through the filter until the test agent is detected in the air sample at the outlet of the filter (breakthrough concentration) or until the filter protection time, which is set by the RPE manufacturer, expires.
Features:
Universal platform for research & quality control of RPE
It is suitable for most tests of filters according to international standards, and in case of changes or new ones, you do not have to partially or completely replace the equipment - the settings are simply and flexibly changed in the software.
Modern way of control
Used a touch-screen display and the app on personal computer with a simple and convenient interface in English.
Increases the productivity of the testing process
Quickly comes to the mode (from 20 to 60 min, depends on test agent).
The tests do not require the constant operator presence. The software automatically maintains test parameters, records and stores results.