The DRXD-001 Online Free Chlorine analyzer (Constant voltage method) consists of a flow pool (embedded electrode module), a constant current system and a transmitter. The above components are jointly installed on the ABS panel, with a high degree of intelligence and flexibility, easy to install and integrate.
Technical specification
Measurement method - constant voltage three electrode measurement, with temperature, pH compensation
Measuring range & accuracy
(Residual chlorine)
0-1mg/L (±0.02mg /L or ±5% MAX),
1-3mg/L (±0.05mg /L or ±5%FS),
3-5mg/L (±0.2mg /L or ±5%FS), .
Compensation range - temperature 0 ~ 40℃ pH 6.0 ~ 8.5
Response time - < 1min, T90
Circulation pool material - organic glass
Working pressure - 0.15 ~ 3bar
Allowable flow rate - 0.5 ~ 2L/min
Ambient temperature - -10 ~ 50℃
Water sample temperature - 5 ~ 45℃
Cleaning function - unique electrochemical + mechanical automatic cleaning
Transmitter
Resolution - 0.01mg/L
Analog output - Optional 2 channels 4 ~ 20mA, loop load 500Ω
Digital output - one RS485 Modbus communication protocol for optional
Control output - Optional 2-way alarm relay and 1-way cleaning relay (250VAC\10A)
Display - 240×160 graphic lattice backlight liquid crystal, with LED
Ambient temperature - -10 ~ 55℃
Relative humidity - 5% ~ 90%, no condensation
Shell material - Polycarbonate (PC)
Power supply - 220VAC, 50Hz, 20W Standard,24VDC±10%, 20W is optional
Installation mode - wall-hanging installation
Size - 234/185/118mm, L/W/H
Weight - 1.5kg