Automatic testing process, put the sample, press the "Start" button to obtain the hardness value of each sample. Samples automatically rise (no height restrictions). Automatic initial test force is applied, the main test force is automatically applied. Automatic measurement of hardness, hardness values of automatic digital display.
Touch screen interface, easy operation, multi-language versions. Available for using after staring up, unnecessary for installing weights and Debugging Electric loading, closed-loop control.
Automatically apply the initial load and main load. Automatic correction of high, medium and low hardness numbers.
Exchanging scales
Rockwell, Brinell, Vickers
Configuration
Z axis rising, X and Y axis sample table moves automatic
Samples rise
Automatic
Automatic sample rate rise
3mm/s
Hardness testing range
HRA:20-88、HRB:20-100 、HRC:20-70 、HRD:40-77 、HRE:70-100 、
HRF:60-100 、HRG:30-94、HRH:80-100、HRK:40-100、HRL:100-120
HRM:85-115、 HRR:114-125
Testing range of Superficial Rockwell
HR15N:70-94 HR30N:42-86 HR45N:20-77 HR15T:67-93 HR30T:
29-82 HR45T:1-72
Rockwell initial test force
10kgf(98.07N),3.0kgf(29.4N)
Rockwell total test force
60kgf(588N),100kgf(980N)
150kgf(1471N),15kgf(147.1N)
,30kgf(249.2N),45kgf(441.3N)
Dwell time
0-99s
Hardness resolution
0.1HR
Test force error
<0.5%
Carried standard
BSEN 6508, ISO 6508, ASTM E18, GB/T230
Max. Height of sample
50mm
The distance of indenter to the wall
200mm