NDIR, CO2

Description

MEMSF MTP40-F NDIR CO2 Sensor Module Target Gas: Carbon Dioxide (CO2) Operating Principle: Non-dispersive infrared (NDIR) Measurement Range: 400ppm--5000 ppm Measurement Interval (s): 2s Accuracy: ±(50ppm+5% of reading) Response Time: T90 time is 90s Operating Temperature: 0-50°C Operating Humidity: 0-90% RH non condensed Storage Temperature: -20°C---60°C Dimensions(mm): 32.5×20×8.5mm (max dimensions) Power Supply: 4.2V--5.5V Power Consumption: 300mA peak, 4mA normal, 13mA average Life Expectancy: 10 years+ Serial Communication: UART/ IIC PWM Output: Period: 1004ms, Pulse: 2ms-1002ms (400--5000ppm) Alarm Output: >1000ppm output to 1, <800ppm output to 0, open-drain output with pull-up resistor, the pin cannot sink any current Self Calibration Cycle: Initial time is 24 hours, later is 7 days