SITRANS SL is designed for measuring oxygen (O2) with high sensitivity. It consists of a pair of cross-duct sensors, a transmitter unit and a detector unit, both with the same dimensions. Due to its little installation effort and its minimum maintenance requirements, SITRANS SL guarantees a high level of economic efficiency.
SITRANS SL – easy to handle with contact-free measurement
The gas analyzer SITRANS SL consists of a pair of cross-duct sensors, a transmitter unit and a detector unit. The hardware processes the measurement signal into a concentration value as well as monitors, controls and servs the communication. The sensors can operate under harsh conditions. Measurements are carried out with a short response time permitting effective cost-savings. The laser characteristics allow single-line spectroscopy free of interferences which means that only the gas component of interest will interact with it and other process influences are eliminated.
High functionality with remote control
The detector unit is connected to the transmitter unit by means of a sensor cable. A further cable on the receiver is used to connect the power supply and the communication interfaces. The receiver enclosure contains a local user interface (LUI) with an LC display which can be read through a window in the cover. The LUI is operated by remote control.
Innovative solution for calibration
An in-line reference cell filled with a non-interfering reference gas, a further development of the already established Siemens TDL technology, provides unique user benefit in terms of minimum need for maintenance and recalibration.
Applications
Process control (chemical industry)