The TBLC08 50uH LISN is a Line Impedance Stabilization Network for the measurement of line-conducted interference within the range of 9kHz to 30MHz, according to the CISPR16 standard.
The device is designed for testing single phase, AC-powered equipment with supply voltages up to maximum 240V.
Conducted noise can be measured on the phase and on the neutral conductor.
The TBLC08 is equipped with a switchable limiter/attenuator and an artificial hand connection.
The device is available with country-specific DUT connectors.
SPECIFICATIONS
Frequency range: 9 kHz to 30 MHz
Impedance: 50 Ω ║ (50 µH + 5 Ω)
Artificial hand: 220 pF + 511 Ω
Switchable PE: 50 Ω ║ 50 µH
Limiter / attenuator: 150 kHz to 30 MHz; 10 dB
Air core inductors
Line voltage: nominal 0-240V; max. 260V DC or AC / 50-60 Hz; CAT II
Max. current: 8A @ 23°C
DUT socket: country specific
Measurement connector: 50 Ω BNC
Power connector: IEC 60320 C13
Operating Temperature Range: +5°C … + 40°C; 5% to 80% RH
Safety: Safety Class I, IEC 1010-01