Designed to detect hydrocarbons (C1 - C9+) which may not absorb in the UV region
Temperature Controlled Stripper System effectively strips hydrocarbons from water by cross flowing N2 over the packed stripper column. Includes Drier Assembly to ensure a dry sample gas is delivered to the analyzer.
Dual-sided, freestanding rack with Stripper Sample System mounted on the backside and Model 402REU NEMA-4 enclosure provided on front
Model 402REU supplied as a 19-inch sheet metal relay rack enclosure with: temperature controller and isothermal chamber; stainless steel sample system including pressure gauges and regulators, integral backpressure regulator; tubing and fittings; integral digital meter; electronic flame guard circuit and fuel shut-off solenoid valve
Flame ionization detector (FID) for measurement of total hydrocarbons at trace levels in a gaseous atmosphere
System NEMA-4 enclosure is provided with a meter viewing window
NEMA-4 enclosure can be X-Purged for hazardous locations (see options)
Standard ranges: 0-10, 0-100, 0-1,000 ppm hydrocarbons (methane equivalent)
Fuel: Mixture of 40% hydrogen, 60% nitrogen
Output: 0-1 VDC and 4-20 mADC, Isolated current output
Sample Selector Module: Integral system for control of sample, span and zero gases. Includes three- way valve, sample bypass flowmeter, needle valve and all necessary tubing and fittings in stainless steel.
X and Z purge options available with additional cost