• Source 4~20mA, 0~24mA with 1|jA resolution (1KQ load, 24V loop supply), basic accuracy 0.025%.
• Source 0~24.000V with lmV resolution. Source frequency l~20000Hz square wave with programmable duty cycle (1~99%). Auto step and auto ramp for sourcing mA, V.
• Easy fixed 25% up and down, or programmable percentage (1~99%) increase and decrease (mA, V).
• Measure DC mA with loop power (24V output).
• Measure DC V.
• Measure Température (°C, °F) of 11 types of thermocouples (K, J, E, T, R, S, N, L, U, B and C).
• Calibrator grade précision température measurement with 0.1*C & 0.1*F resolution.
• Détection of thermocouples disconnection.
• Source and Measure at the same time. Warning for overload, output open (mA) or short (V).
• Short circuit protection for output.
• Battery power indication (%) when turn on the calibrator.
• Calibration of 4~20mA panel meters.
• LED brightness testing (0~24mA).
• Température calibration of panel meters or instruments for 11 types of thermocouples.